Optimizing brain tissue contrast with EPI: a simulated annealing approach.
Ikonomidou VN, van Gelderen P, de Zwart JA, Fukunaga M, Duyn JH.
Advanced MRI Section, LFMI, NINDS, National Institutes of Health, Bethesda, Maryland 20892-1065, USA. viko@nih.gov
A new magnetization preparation and image acquisition scheme was developed to obtain
high-resolution brain images with optimal tissue contrast. The pulse sequence was
derived from an optimization process using simulated annealing, without prior
assumptions with regard to the number of radiofrequency (RF) pulses and flip
angles. The resulting scheme combined two inversion pulses with the acquisition
of three images with varying contrast. The combination of the three images
allowed separation of gray matter (GM), white matter (WM), and cerebrospinal
fluid (CSF) based on T1, contrast. It also enabled the correction of small
errors in the initial T1 estimates in postprocessing. The use of
three-dimensional (3D) sensitivity-encoded (SENSE) echo-planar imaging (EPI)
for image acquisition made it possible to achieve a 1.15(3) mm3 isotropic
resolution within a scan time of 10 min 21 s. The cortical GM signal-to-noise
ratio (SNR) in the calculated GM-only image varied between 30 and 100.
The novel technique was evaluated in combination with blood oxygen level-dependent
(BOLD) functional magnetic resonance imaging (fMRI) on human subjects, and provided
for excellent coregistration of anatomical and functional data. 2005 Wiley-Liss,
Inc
PMID: 16032676 [PubMed - indexed for MEDLINE]