Cartilage volume quantification via Live Wire segmentation.
Gougoutas AJ, Wheaton AJ, Borthakur A, Shapiro EM, Kneeland JB, Udupa JK, Reddy R.
Metabolic Magnetic Resonance Research and Computing Center, Department of Radiology, University of Pennsylvania,
B1 Stellar-Chance Laboratories, 422 Curie Blvd, Philadelphia, PA 19104-2045, USA.
RATIONALE AND OBJECTIVES: A reduction in cartilage volume is characteristic of osteoarthritis
and hence there exists a need for an accurate and reproducible method to measure in vivo
cartilage volume. Quantification of cartilage volume from magnetic resonance (MR) images
requires a segmentation technique such as the user-driven "Live Wire" strategy that can
reliably delineate object volumes in a time-efficient manner. In the present work, the
accuracy and reproducibility of the Live Wire method for the quantification of cartilage
volume in MR images is evaluated. MATERIALS AND METHODS: The accuracy of the Live Wire
method was assessed by comparing the MR-based volume measurement of a patellar cartilage-shaped
phantom versus data calculated via water displacement. The inter- and intra-operator
reproducibility of the technique was evaluated from Live Wire segmentation of the patellar
cartilage volume from fat-suppressed 3-dimensional spoiled-gradient-echo images of five
healthy human volunteers performed by three operators. To provide data for analysis of
inter-scan reproducibility, the human scans were repeated five times with the aid of a
leg-restraining jig to minimize repositioning error. RESULTS: The volume of the patellar
cartilage-shaped phantom measured via Live Wire segmentation of MR images was within 97.8%
of its true volume. The average inter- and intra-operator coefficients of variation of
three operators were 3.0% and 0.4%, respectively. The average inter-scan coefficient of
variation of five repeated scans of each volunteer was 2.7%. CONCLUSION: The data suggest
that the Live Wire strategy is an accurate, reproducible, and efficient technique to measure
cartilage volume in vivo in a feasible amount of operator time.
PMID: 15596377 [PubMed - indexed for MEDLINE]